Early loosening and secondary dislocation due to a broken trochanteric osteotomy wire following a Charnley total hip arthroplasty: a case report
Cases Journal volume 2, Article number: 7117 (2009)
Abstract
We report a case of interposition of a broken trochanteric wire in the hip joint. This caused early wear of the prosthesis and dislocation of the Charnley total hip arthroplasty. The patient was treated with a revision total hip arthroplasty. This rare complication should be taken into consideration when performing a trochanteric osteotomy fixation with wiring in Charnley total hip arthroplasty.
Introduction
Trochanteric osteotomy is a known method used in Charnley total hip arthroplasty. The advantage of the transtrochanteric approach is that it gives a better view of the hip joint. The disadvantages of this method are increased operative time, increased blood loss, trochanteric non union and trochanteric bursitis [1, 2]. We report a rare complication associated with trochanteric osteotomy.
Case presentation
A 61 year old British white male was admitted with a dislocation of a left Charnley total hip arthroplasty which was performed in 2005 for a subcapital fracture of the left neck of femur. The operation was successful for three years and the patient was doing well. Immediate post operative plain X-ray of the pelvis was satisfactory (Figure 1). In March 2008 the patient presented with a sudden left hip pain while was getting into bed and was unable to weight bear afterwards. On examination, the left leg was shortened and externally rotated.
Plain X-ray of the pelvis showed a dislocated left Charnley total hip arthroplasty and it also showed a foreign body in the hip joint (broken wire) (Figure 2). The hip was reduced under general anaesthesia. The hip joint was stable up to 90 degrees flexion and 10 degrees of internal rotation.
In April 2008, the patient had a revision of left Charnley total hip arthroplasty with a good result. Intraoperatively, a broken wire was found embedded in the polyethylene acetabular component and the polyethylene bearing surface was found to be worn (Figure 3).
Discussion
Articular interposition of a broken wire in the hip joint is a rare complication following Charnley total hip arthroplasty. Four cases have been reported in the literature. The first case was a 53 year old male who had a revision arthroplasty because of a worn acetabular cup due to an intra-articular wire fragments found two years after Charnley total hip arthroplasty [3].
The second case was a 64 year old male who was found to have a wire on the X-rays after three years of the operation. He did not have any symptoms but after 12 years required revision arthroplasty due to cup wear [4].
The third case was an 80 year old female who experienced multiple dislocations of the hip 10 years following Charnley total hip arthroplasty. The wires were taken out arthroscopically [5].
The fourth case was an 80 year old female who had recurrent dislocations 10 years after the operation. Plain X-rays showed intra-articular wires. She required a revision arthroplasty due to prosthetic components wear [6].
In our case, the patient did not have any problems post operatively for three years until he experienced his first dislocation.
Conclusion
The interposition of broken trochanteric wires in the hip joint post Charnley total hip arthroplasty can cause a third body effect leading to early loosening and early wear of the prosthetic components leading to dislocation of the hip. It is a complication that should be considered when performing a Charnley total hip arthroplasty which might necessitate an early revision surgery.
Consent
The patient agreed for this case report to be published after obtaining informed consent. However, the form at that time was not available to be signed.
References
Wroblewski B: Revision Surgery in Total Hip Arthroplasty. 1990, Springer- Verlag, 19-28. 1
Charnley J, Ferreira A: Transplantation of the greater trochanter in arthroplasty of the hip. J Bone Joint Surg. 1994, 46B: 191-197.
Bronson J: Articular interposition of trochanteric wires in a failed total hip replacement. Clin Orthop. 1986, 121: 50-52.
Ritter M, Meding J: Intra-articular trochanteric wire migration following bilateral total hip arthroplasty. Orthopaedic. 1998, 11: 1295-1297.
Lim A, Parekh S, Smith D: Intra-articular migration of broken trochanteric wires. Injury. 1993, 24: 201-202. 10.1016/0020-1383(93)90299-L.
Fakih R, Treacy RBC: Articular Interposition of Broken Trochanteric Wires. Bulletin Hospital for Joint Diseases. 1998, 57: 108-110.
Author information
Authors and Affiliations
Corresponding author
Additional information
Competing interests
The authors declare that they have no competing interest.
Authors' contributions
YS wrote the first draft and RC revised it. SA also revised it and added comments. MM is the senior author and he revised the final draft before submission.
Rights and permissions
Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Shahin, Y., Choudhary, R., Al-Naser, S. et al. Early loosening and secondary dislocation due to a broken trochanteric osteotomy wire following a Charnley total hip arthroplasty: a case report. Cases Journal 2, 7117 (2009). https://doi.org/10.1186/1757-1626-2-7117
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/1757-1626-2-7117