Intrathoracic migration of a Steinman wire: a case report and review of the literature

Abstract

Introduction

Migration of orthopaedic fixation wires into the thoracic cavity occurs infrequently, but can have dire consequences. Although rare, intrathoracic migration is a serious complication that demands immediate removal.

Case presentation

We present a case of a 59-year-old man with an intrathoracic migration of a Steinman wire used for the treatment of a shoulder fracture. Surprisingly, the migration was asymptomatic. The Steinman wire was successfully retrieved from the thorax via thoracotomy.

Conclusion

The migration of pins and wires can cause fatal complications and should be considered as very hazardous. Therefore, if wires need to be used, terminally threaded pins are safer and the free end should be bent. The patients should be frequently followed, both clinically and radiographically, until all the wires are removed.

Introduction

Pins and wires are used extensively for internal fixation of fractures in orthopaedic surgery, especially in the management of fractures and dislocations of the shoulder. Some of these internal fixation devices (Steinman, Kirschner wires) have a tendency to migrate. Although uncommon, when the migration of these wires occurs, there could be devastating consequences. Thus, prompt recognition and immediate retrieval of the implant is paramount to avert those complications [1, 2]. In this study, we present the case of a 59-year-old man in whom a threaded Steinman, previously placed into the right shoulder, fixing a fracture, migrated into the right lung parenchyma; and was successfully retrieved via thoracotomy.

Case presentation

In September 2007, a 59-year-old Caucasian, drug addicted man was admitted into the orthopaedic department of our clinic after a low-velocity, three-part fracture of the right humerus head (Neer type IV - surgical neck and greater tuberosity fragment). He had a medical history of HBV, HCV and HIV positive, Brody myopathy and excised eosinophil granule of the 3rd and 4th right ribs. After a closed reduction, three percutaneous Steinman wires (threaded) were used for the fixation of the fracture (Figure 1). The patient was discharged the second postoperative day. On the third follow-up, three weeks postoperative, radiography was performed and surprisingly one Steinman was noted to be missing from the right shoulder. A chest radiograph confirmed its migration into the right hemithorax (Figure 2). The patient had no cardiopulmonary symptoms, and there was no evidence of a pneumothorax or a hemothorax. A chest CT-scan was performed (Figure 3) and the wire was located in the superior segment of the right lower lobe, but did not enter the mediastinum or the pleural space.

Figure 1
figure 1

Post-operative radiograph of the three-part fracture of the right humerus head be fixed with three percutaneous Steinman wires.

Figure 2
figure 2

Intrathoracic migration of a Steinman wire; post-operative radiograph on the third week.

Figure 3
figure 3

A chest CT-scan that performed, presented the Steinman wire to be located in the superior segment of the right lower lobe.

Because of the central position of the pin, prompt removal was recommended. The patient was scheduled for surgery the following day. He was placed in a left lateral decubitus position and a posterolateral thoracotomy in the 5th right intercostal space was used. The Steinman wire, be lodged in the parenchyma of the right lower lobe, was grasped and slowly withdrawn from the lung. Once removed, the lung was re-expanded without difficulty. As a precaution, a 28F chest tube was placed under direct vision. There was no air leak or bleeding postoperatively and the tube was removed on the second postoperative day. The patient was discharged the following day. Four weeks later the other two Steinman wires on the right humeral head were removed under local anaesthesia.

Discussion

Orthopaedic procedures frequently involve utilization of internal fixation devices (Steinman, Kirschner wires and pins). Migration of these devices within the chest is uncommon, but is a known complication, particularly with fixation wires or pins around the shoulder. Mazet reported in 1943 the first two cases of migration of a Kirschner wire from the shoulder region [1]. In 1990, Lyons and Rockwood published a review of all the reported cases of pin migration in operations on the shoulder [2]. It is noteworthy that, in 17 out of 47 cases the wire migrated to a major vascular structure, and in 21 out of 47 the wire had been used for internal fixation of a dislocated sternoclavicular joint [2].

Migrations usually follow a retrograde path and the wires protrude near the entry point. Occasionally migration occurs in antegrade direction, as in our case [3]. A recent review of the literature showed sporadic case reports of migrated pins and wires from the shoulder region to the spinal canal, the trachea, the spleen, the pulmonary artery, the heart, the mediastinum, the lung, the subclavian artery and the ascending and abdominal aorta [4]-[6]. Migration of K-wires has been reported as early as the first day and as late as 21 years [7] after the fixation. Usually the process causes no symptoms, but there have been observed deaths attributable to migration. All deaths were associated with catastrophic cardiovascular events and cardiac tamponade. Serious, non-fatal complications include pericardial tamponade, arrhythmia, pericarditis, pseudo-aneurysm, aortopulmonary fistula, pneumothorax, haemoptysis, subclavian steal syndrome, hemianopia, hemiplegia, paraplegia, radicular pain, dysphagia and splenic hematoma [4, 6].

Various theories have been proposed to explain this migration, including muscular activity, regional resorption of bone, and the great freedom of motion of the shoulder. The latter may be responsible for the breakage of pins and wires that have been reported. The migration to the thorax seems to be favored by respiratory excursions, negative intrathoracic pressure, and gravitational forces [4, 8]. This severe complication can be theoretical reduced by using threaded pins in surgical operations of the shoulder, by bending the free end of the wire, and lastly by removing the hardware as soon as possible [2, 3]. Notwithstanding, in practice all these does not fully, one hundred percent, protect the patient against pin migration.

However, as evident in Chang et al. case, fracture and migration of a fractured fragment can still occur even if the outer end of the pin is bent and well-fixed [8]. Therefore, regular radiographic follow-up is needed for patients with internal fixation devices; any fractured pins or wires must be removed immediately to prevent dangerous migration. In other words, patients should be frequently followed both clinically and radiographically [4]. CT scan clearly evaluates the exact location of the migrated pin or wire, as well as existing complications.

Once migration of a pin or K-wire or Steinman is recognized, immediate surgical removal should be carried out. Sternotomy, thoracotomy and video-assisted thoracoscopy have been used [8]. Thoracoscopic removal of intrathoracic foreign bodies can be accomplished safely if the object can be withdrawn through a port site, if it does not traverse the mediastinum, and if the patient can tolerate single-lung ventilation [9]. Otherwise, thoracotomy and sternotomy are safe and could be used to remove migrated wires. Sternotomy may be considered more appropriate in the presence of injury to the heart or great vessels [5, 10].

Conclusion

Consequently, although our patient underwent a successful removal of the wires and recovered uneventfully, the migration of pins and wires can cause fatal complications and should be considered as very hazardous. Therefore, if wires need to be used, terminally threaded pins are safer and the free end should be bent. The patients should be frequently followed, both clinically and radiographically, until all the wires are removed.

Consent

Written informed consent was obtained from the patient for publication of this case report and accompanying images. A copy of the written consent is available for review by the Editor-in-Chief of this journal.

References

  1. Mazet RJ: Migration of a Kirschner Wire from the Shoulder Region into the Lung. Report of Two Cases. J Bone Joint Surg. 1943, 25: 477-483.

    Google Scholar 

  2. Lyons FA, Rockwood CA: Migration of pins used in operations on the shoulder. J Bone Joint Surg. 1990, 72: 1262-1267.

    CAS  PubMed  Google Scholar 

  3. Marya KM, Yadav V, Rattan KN, Kundu ZS, Sangwan SS: Unusual K-wire Migration. Indian J Pediatr. 2006, 73: 1107-1108. 10.1007/BF02763056.

    Article  CAS  PubMed  Google Scholar 

  4. Sarper A, Urgüden M, Dertsiz L, Demircan A: Intrathoracic migration of Steinman wire. Interact Cardiovasc Thorac Surg. 2003, 2: 210-211. 10.1016/S1569-9293(03)00047-1.

    Article  PubMed  Google Scholar 

  5. Venissac N, Alifano M, Dahan M, Mouroux J: Intrathoracic migration of Kirschner pins. Ann Thorac Surg. 2000, 69: 1953-1955. 10.1016/S0003-4975(00)01198-X.

    Article  CAS  PubMed  Google Scholar 

  6. Wada S, Noguchi T, Hashimoto T, Uchida Y, Kawahara K: Successful migration of a patient with penetrating injury of the esophagus and brachiocephalic artery due to migration of Kirschner wires. Ann Thorac Cardiovasc Surg. 2005, 11: 313-315.

    PubMed  Google Scholar 

  7. Kumar P, Godbole R, Rees GM, Sarkar P: Intrathoracic migration of a Kirschner wire. J R Soc Med. 2002, 95: 198-199. 10.1258/jrsm.95.4.198.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Chang HC, Lim KE, Hsu YY, Tsai YH, Chan YI, Kuo HW: Pneumothorax due to Migration of a Fractured Kirschner Wire - A Case Report. Tzu Chi Med J. 2007, 19: 87-89.

    Article  Google Scholar 

  9. Calkins CM, Moore EE, Johnson JL, Smith WR: Removal of an Intrathoracic migrated fixation pin by thoracoscopy. Ann Thorac Surg. 2001, 71: 368-370. 10.1016/S0003-4975(00)01722-7.

    Article  CAS  PubMed  Google Scholar 

  10. Marchi E, Reis MP, Carvalho MV: Transmediastinal migration of Kirschner wire. Interact Cardiovasc Thorac Surg. 2008, 7: 869-870. 10.1510/icvts.2008.185850.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Dimitrios D Nikolopoulos.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

DN conducted a literature search, prepared the final manuscript and operated the patient. NS conducted a literature search, contributed to the preparation of the manuscript and operated the patient. DY prepared the first draft of the manuscript and treated the patient. EV and MK supervised the manuscript and treated the patient. All authors participated in the surgical operation, and contributed equally in the post-operative patient's follow-up. All authors contributed in collecting patient data and editing arthroscopic operative images.

Rights and permissions

Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Sergides, N.N., Nikolopoulos, D.D., Yfadopoulos, D.K. et al. Intrathoracic migration of a Steinman wire: a case report and review of the literature. Cases Journal 2, 8321 (2009). https://doi.org/10.4076/1757-1626-2-8321

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.4076/1757-1626-2-8321

Keywords